10 Channel

10 Channel Detector circuit overlay

10 Channel Detector circuit overlay

10 Channel Thermopile Detector

10 Channel Thermopile Detector

Features
A ten-channel thin-film thermopile in a TO-8 package. Each active area is 3.16mm x 0.4mm and offers low noise voltage. Internal aperture minimizes channel-to-channel crosstalk while increasing sensitivity.

Applications
The perfect single package choice for analyzing multiple anesthetic gases in demanding environments such as hospital operating rooms. Potential applications in non-contact temperature and spectral discrimination of materials.

Benefit
Ten-channel spectral differentiation for demanding applications.

Options

  1. See Standard Windows and Filters (PDF) for list of optical filter options
  2. Internal 30kΩ 5% NTC chip thermistor provides ambient package temperature measurement. See DC-4005 section of Thermistor Options (PDF).
  3. See Thermopile Configuration Table for more options.

Datasheet
For more detailed information see the datasheet 10 Channel (PDF).

Parameter Min Typical Max Symbol Units Comments
Active Area size   3.16 x .4   AA mm Hot junction size, per element.
Element Area   1.264   A mm2  
Number of Junctions   40       Per element.
Number of Channels   10       Per detector package.
Output Voltage 90 115 130 Vs μV DC,  H=330μW/cm2 (3)
Signal-to-Noise Ratio 6,429 10,088 16,049 SNR √Hz DC, SNR=Vs/Vn
Responsivity 21.6 27.6 31.2 V/W DC, =Vs/HA  (2)
Resistance 4.0 8.0 12 R Detector element
Temperature Coefficient of   -.36     %/°C Best linear fit, 0° to 85°C (1)
Temperature Coefficient of  R   -.2     %/°C Best fit, 0° to 85°C (1)
Noise Voltage 8.1 11.4 14.0 Vn nV/√Hz Vn2=4kTR
Noise Equivalent Power .26 .42 .65 NEP nW/√Hz DC,  NEP= VnHA/Vs (2)
Detectivity 1.7 2.7 4.3 D* 108√Hz/W DC,  D*=Vs/ VnH√A  (2)
Time Constant   38   Τ ms Chopped, -3dB point (1)
Field of View   N/A   FOV Degrees See Assembly Drawings for FOV Description.
Package Type   TO-8 w/12 Pins       Standard package hole size:Ø.437
Element Matching   10 20 M % M=|VA-VB|/VB (2)
Element Separation   1.8 & 4.28     mm Center to Center
Operating Temperature -50   100 Ta °C  

General Specifications: 
Flat spectral response from 100nm to > 100µm.  Linear signal output from 10-6 to 0.1W/cm2
Maximum incident radiance 0.1W/cm2 , damage threshold ≥ .5W/cm2
Notes:      
(1)
Parameter is not 100% tested. 90% of all units meet these specifications.
(2)
A is detector area in cm2.
(3)
Test Conditions: 500K Blackbody source; Detector active surface 10cm from 0.6513cm Diameter Blackbody Aperture.

Product Finder