S60M TO-5

S60M Detector circuit overlay

S60M Detector circuit overlay

S60M TO-5 Thermopile Detector

S60M TO-5 Thermopile Detector

Features
A single-channel silicon-based thermopile with very high output for its small 0.6mm x 0.6mm active area in a TO-5 package. Delivers a very low Temperature Coefficient of Responsivity of -0.04%/°C.

Applications
High output, small active area, fast time constant that has a higher cost.

Benefit
Excellent for non-contact temperature and gas analysis.

Options

  1. See Standard Windows and Filters (PDF) for list of optical filter options
  2. Internal aperture precisely defines active area for applications with FOV and/or spot size requirements. See Aperture Options (PDF) for available sizes.
  3. See Thermopile Configuration Table for more options.

Datasheet
For more detailed information see the datasheet S60M TO-5 (PDF).

Specifications apply at 23°C with KBr Window and Argon encapsulating gas.

Parameter Min Typical Max Symbol Units Comments
Active Area size   .6 x .6   AA mm Hot junction size, per element.
Element Area   .36   A mm2  
Number of Junctions   72       Per element.
Number of Channels   1       Per detector package.
Output Voltage 90 120 140 Vs μV DC,  H=330μW/cm2 (3)
Signal-to-Noise Ratio 2,123 3,125 4,294 SNR √Hz DC, SNR=Vs/Vn
Responsivity 75.8 101.0 117.8 V/W DC, =Vs/HA  (2)
Resistance 65 90 120 R Detector element
Temperature Coefficient of   -.04     %/°C Best linear fit, 0° to 85°C (1)
Temperature Coefficient of  R   .11     %/°C Best fit, 0° to 85°C (1)
Noise Voltage 32.6 38.4 42.4 Vn nV/√Hz Vn2=4kTR
Noise Equivalent Power .28 .38 .56 NEP nW/√Hz DC,  NEP= VnHA/Vs (2)
Detectivity 1.1 1.6 2.2 D* 108√Hz/W DC,  D*=Vs/ VnH√A  (2)
Time Constant   27   Τ ms Chopped, -3dB point (1)
Field of View   64°/81°   FOV Degrees See Assembly Drawings for FOV Description.
Package Type   TO-5       Standard package hole size:Ø.150"
Operating Temperature -50   100 Ta °C  

General Specifications: 
Flat spectral response from 100nm to > 100µm.  Linear signal output from 10-6 to 0.1W/cm2
Maximum incident radiance 0.1W/cm2 , damage threshold ≥ .5W/cm2

Notes:      
(1) Parameter is not 100% tested. 90% of all units meet these specifications.
(2) A is detector area in cm2.
(3)Test Conditions: 500K Blackbody source; Detector active surface 10cm from 0.6513cm Diameter Blackbody Aperture.

 

* Protected by U.S. Patent No. 5, 059,543 and U.S. Patent No. 5,100,479