TM34

TM34 Detector circuit overlay

TM34 Detector circuit overlay

Features
A three-channel thin-film thermopile in a TO-8 package. Each symmetrically positioned active area is 3.16mm x 0.4mm. Offers low noise output and internal aperture minimizing channel-to-channel crosstalk while increasing sensitivity.

Applications
Gas analysis for automotive, environmental air quality, industrial sensors and medical monitoring.

Benefit
Low noise and 3-channels in a TO-8 package with moderate output.

Options

  1. See Standard Windows and Filters (PDF) for list of optical filter options
  2. Internal 30kΩ 5% NTC chip thermistor provides ambient package temperature measurement. See DC-4005 section of Thermistor Options (PDF).
  3. See Thermopile Configuration Table for more options.

Datasheet
For more detailed information see the datasheet TM34 (PDF).

Parameter Min Typical Max Symbol Units Comments
Active Area size   3.16 x .4   AA mm Hot junction size, per element.
Element Area   1.264   A mm2  
Number of Junctions   40       Per element.
Number of Channels   3       Per detector package.
Output Voltage 90 115 130 Vs μV DC,  H=330μW/cm2 (3)
Signal-to-Noise Ratio 6,429 10,088 16,049 SNR √Hz DC, SNR=Vs/Vn
Responsivity 21.6 27.6 31.2 V/W DC, =Vs/HA  (2)
Resistance 4.0 8.0 12 R Detector element
Temperature Coefficient of   -.36     %/°C Best linear fit, 0° to 85°C (1)
Temperature Coefficient of  R   -.2     %/°C Best fit, 0° to 85°C (1)
Noise Voltage 8.1 11.4 14.0 Vn nV/√Hz Vn2=4kTR
Noise Equivalent Power .26 .42 .65 NEP nW/√Hz DC,  NEP= VnHA/Vs (2)
Detectivity 1.7 2.7 4.3 D* 108√Hz/W DC,  D*=Vs/ VnH√A  (2)
Time Constant   38   Τ ms Chopped, -3dB point (1)
Field of View   NA   FOV Degrees See Assembly Drawings for FOV Description.
Package Type   TO-8 with 3 Pins       Standard package hole size:Ø.150"
Operating Temperature -50   100 Ta °C  

General Specifications: 
Flat spectral response from 100nm to > 100µm.  Linear signal output from 10-6 to 0.1W/cm2
Maximum incident radiance 0.1W/cm2 , damage threshold ≥ .5W/cm2

Notes:      
(1) Parameter is not 100% tested. 90% of all units meet these specifications.
(2) A is detector area in cm2.
(3)Test Conditions: 500K Blackbody source; Detector active surface 10cm from 0.6513cm Diameter Blackbody Aperture.

Product Finder